pHotoFlex
®
STD
Operation
53
ba75958d05
10/2017
Socket assignment
4.7.2
Configuring the RS232 interface
For an error-free data transmission, the RS232 interface should be set
to the same transmission speed (
Baud rate
) for the pHotoFlex
®
STD
and PC/printer.
You can set the following values for the baud rate on the
pHotoFlex
®
STD : 1200, 2400, 4800, 9600, 19200.
The baud rate is selected in the menu,
Configuration
/
System
/
Inter-
face
/
Baud rate
.
4.7.3
Selecting the output format of datasets
For downloading data to the interface you can select the output format.
It is selected in the menu,
Configuration
/
System
/
Interface
/
Output
format
.
The ASCII output format delivers formatted datasets.
The CSV output format delivers datasets separated by ";".
Output format, ASCII
RS 232
REC
1
3
4
5
6
2
1 RTS
2 RxD
3 TxD
4 DTR
5 SG
6 CTS
pHotoFlex
Ser. no.
12345678
01.12.11 09:56:20
ID 1/ PROG 83
1.74 mg/l NH4-N [BV]
________________________________________
03.12.11 14:01:57
ID 2/ PROG 81
176 mg/l O2 [BV]
________________________________________
etc...