pHotoFlex
®
STD
Operation
35
ba75958d05
10/2017
4.5.3
Measuring the concentration
From the second concentration measurement, the data of the program
last used is automatically displayed here.
With
<
> <
>
you can quickly switch between the ten programs last
used.
To select a program, you can as an alternative also read in the program
number of an analysis specification with a bar code reader (see
section 8.2). The following step three is skipped. You can directly start
measuring.
The program number of the test is given in the analysis specification, in
the list of available programs and, with some tests, on the packing of
the test (under the bar code).
If a program number is selected that requires a measured blank value,
the menu automatically guides to the blank value measurement.
1
If necessary, press the
<M>
key several times until the mea-
sured parameter
Concentration
is selected.
2
Open the
Program number
display with
<PROG>
,
enter the required program number with the number keys and
confirm with
<START/ENTER>
.
or (from the second concentration measurement):
Select a program out of the last ten programs with
<
> <
>
.
The program data is displayed.
Photometry
\
Concentration
i
Select program
with <PROG>
01.12.11 15:12
Photometry
\
Concentration
i
Select program
with <PROG>
or with
?
?
83: A6/25 MC
NH4-N
16 mm
0.20 - 8.00 mg/l
01.12.11 15:12
First concentration measure-
ment with the pHotoFlex
®
STD
Second and all further concen-
tration measurements
Summary of Contents for pHotoFlex STD
Page 1: ...OPERATING MANUAL ba75958d05 10 2017 pHotoFlex STD LED FILTER PHOTOMETER ...
Page 6: ...Contents pHotoFlex STD 6 ba75958d05 10 2017 ...
Page 16: ...Commissioning pHotoFlex STD 16 ba75958d05 10 2017 ...
Page 62: ...Operation pHotoFlex STD 62 ba75958d05 10 2017 ...
Page 74: ...Accessories and options pHotoFlex STD 74 ba75958d05 10 2017 ...
Page 78: ...Index pHotoFlex STD 78 ba75958d05 10 2017 Z Zero adjustment 43 ...