pHotoFlex
®
STD
Operation
49
ba75958d05
10/2017
4.6.1
Storing measurement datasets
Proceed as follows to transmit to the data memory and simultaneously
output to the interface a measurement dataset:
A measurement dataset is stored quickly by twice pressing
<STO>
. It
is stored with the ID last set.
If the memory is full
You can erase the entire memory (see section 4.6.5), or overwrite the
oldest dataset with the next storing procedure.
A security prompt appears before a dataset is overwritten.
1
Press the
<STO>
key.
The
Store
display appears.
2
Using
<
> <
>
,
<START/ENTER>
and the number keys,
change and confirm the ID number (
ID
) as necessary (0 ...
999).
3
Using
<START/ENTER>
or
<STO>
, confirm
Store
.
The dataset is stored. The meter switches to the measured val-
ue display.
i
01.12.2011 15:10:11
PROG
86
1.12 mg/l NH4-N
[
BV
]
assign ID
:
1
Store
(ID: 1:
Store
(
996
free
)