pHotoFlex
®
STD
Operation
39
ba75958d05
10/2017
5
Insert a cell with blank sample (see section 4.2).
6
Start the measurement of the blank value with
<START/EN-
TER>
.
After measuring, the result of the blank value measurement is
displayed and stored.
The result is displayed as
i
Blank value measurement suc-
cessful!
or
i
Blank value measurement erroneous!
.
7
Confirm the result with
<START/ENTER>
.
The blank value measurement is completed.
The meter is ready to measure.
or:
Discard the result with
<ESC>
.
Subsequently, carry out a new blank value measurement.
i
83: A6/25 MC
Cell
i
= 16 mm
i
Insert blank sample
Start measurement
Measure blank value
i
83: A6/25 MC
Cell
i
= 16 mm
i
Absorbance
= 0.301
i
Blank value mea-
surement
successful!
Measure blank value