Operation
pHotoFlex
®
STD
52
ba75958d05
10/2017
4.6.5
Erasing stored measurement datasets
You can erase the stored measurement datasets altogether if you no
longer need them.
Erasing all measurement datasets is done in the menu,
Configuration
/
System
/
Measured value memory
/
Delete
.
Erasing individual datasets is not possible. If all memory locations are
occupied, however, it is possible to overwrite the oldest dataset at a
time. A security prompt appears before a dataset is overwritten.
4.7
Transmitting data (RS 232 interface)
Via the RS 232 interface, you can transmit data to a PC or an external
printer.
4.7.1
Connecting a PC/external printer
Connect the interface to the devices via the AK540/B (PC) or AK540/S
(external printer) cable.
The RS232 interface is not galvanically isolated.
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result.
Set up the following transmission data on the PC/printer:
Baud rate
can be selected from: 1200, 2400, 4800, 9600,
19200
The baud rate must agree with the baud rate set on
the PC/printer.
Handshake
RTS/CTS
PC only:
Parity
none
Data bits
8
Stop bits
1s
Summary of Contents for pHotoFlex STD
Page 1: ...OPERATING MANUAL ba75958d05 10 2017 pHotoFlex STD LED FILTER PHOTOMETER ...
Page 6: ...Contents pHotoFlex STD 6 ba75958d05 10 2017 ...
Page 16: ...Commissioning pHotoFlex STD 16 ba75958d05 10 2017 ...
Page 62: ...Operation pHotoFlex STD 62 ba75958d05 10 2017 ...
Page 74: ...Accessories and options pHotoFlex STD 74 ba75958d05 10 2017 ...
Page 78: ...Index pHotoFlex STD 78 ba75958d05 10 2017 Z Zero adjustment 43 ...