Oxi 7310
Data storage
ba75931e03
12/2011
37
Settings
You can configure the automatic storage function with the following settings:
Starting the
automatic storage
function
To start the automatic storage function, select
Continue
with
<
><
>
and
confirm with
<ENTER>
. The meter switches to the measured value display.
The active automatic storage function can be recognized by the progress bar
in the function display. The progress bar indicates the remaining storage dura-
tion.
Menu item
Possible
setting
Description
ID number
1 ... 10000
ID number for the dataset series.
Interval
1 s, 5 s, 10 s,
30 s, 1 min,
5 min,
10 min,
15 min,
30 min,
60 min
Storage interval.
The lower limit of the storage interval
can be restricted by the number of free
storage locations.
The upper limit is restricted by the stor-
age duration.
Duration
1 min ... x min
Storage duration.
Specifies after which time the automatic
storage should be terminated.
The lower limit of the storage duration is
restricted by the storage interval.
The upper limit is restricted by the num-
ber of free storage locations.
1 Remaining storage duration
2 Graphical display of the storage duration
If the automatic storage function is activated, only the following
keys are active: Softkeys,
<M>
,
<STO_>
and
<On/Off>
.
The other keys and the automatic switch-off function are deacti-
vated.
1
2
O
2
7.92
mg/l
25.0
°C
0d03h00min
[CellOx]
01.07.2011
08:00
USB output
Eco mode