Dissolved oxygen
Oxi 7310
24
ba75931e03
12/2011
The minimum duration until a measured value is assessed as stable is the mon-
itored time interval. The actual duration is mostly longer.
5.1.2
Measuring the temperature
For reproducible D.O. measurements, it is essential to measure the tempera-
ture of the test sample.
The temperature is automatically measured by the temperature sensor
(NTC30 or Pt1000) integrated in the sensor.
Measured
parameter
Time interval
Stability in the time inter-
val
D.O. concen-
tration
10 seconds
Δ
: better than 0.05 mg/l
D.O. satura-
tion
10 seconds
Δ
: better than 0.6 %
D.O. partial
pressure
10 seconds
Δ
: Better than 1.2 mbar
The CellOx 325 / StirrOx
®
G and DurOx 325 D.O. sensors have an
integrated temperature sensor.