What to do if...
Oxi 7310
48
ba75931e03
12/2011
11
What to do if...
11.1
Dissolved oxygen
Error message,
OFL, UFL
The measured value is outside the measuring range.
Error message,
Error
Error message,
Leak
11.2
General information
Sensor symbol
flashes
Display
More information and instructions on cleaning and exchange of
sensors are given in the documentation of your sensor.
Cause
Remedy
– Measured value outside the mea-
suring range of the meter
– Use a suitable D.O. sensor
Cause
Remedy
– D.O. sensor contaminated
– Electrolyte depleted
– Clean the D.O. sensor.
– Changing the electrolyte and
membrane cap
– If necessary, replace the D.O.
sensor
Cause
Remedy
– Membrane cap not screwed on
tight enough
– Hole in the membrane
– Screw membrane head tighter
– Replace and refill the mem-
brane cap
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– Batteries almost empty
– Replace the batteries (see