12.1
Measuring ranges, resolution, accuracy
Measuring ranges,
resolution
The values quoted in brackets apply especially to the DurOx 325 sen-
sor.
Parameter
Measuring range
Resolution
D.O. concentration
[mg/l]
0 ... 20.00 (0 ... 20.0)
0 ... 90.0 (0 ... 90)
0.01 (0.1)
0.1 (1)
Saturation [%]
0 ... 200.0 (0 ... 200)
0 ... 600
0.1 (1)
1
D.O. partial pressure
[mbar]
0 ... 200.0 (0 ... 200)
0 ... 1250
0.1 (1)
1
T [°C]
0 ... 50.0
0.1
Accuracy
(± 1 digit)
Parameter
Accuracy
D.O. concentration
[mg/l]
± 0.5 % of measured value
at ambient tempe5 °C ... +30 °C
D.O. Saturation [%]
± 0.5% of measured value
when measuring in the range of ± 10 K
around the calibration temperature
D.O. partial pressure
[mbar]
± 0.5 % of measured value
at ambient tempe5 °C ... +30 °C
T [°C] / temperature sensor
NTC 30
± 0.1
PT 1000
± 0.1
Correction functions
Temperature
compensation
Accuracy better than 2 % at 0 ... +40 °C
Salinity correction
0 ... 70.0 SAL
Air pressure correc-
tion
Automatic through integrated pressure sen-
sor in the range of 500 ... 1100 mbar
The accuracy values specified here apply exclusively to the meter.