Data storage
Oxi 7310
38
ba75931e03
12/2011
Energy saving mode
[Eco mode]
When the automatic storing function is active, the meter provides an energy
saving mode [Eco mode] to avoid unnecessary energy consumption. The
energy saving mode switches off functions of the meter that are not required for
the automatic storage of measurement data (such as the display). By pressing
any key the energy saving mode is switched off again.
Terminating the
automatic storage
function prema-
turely
Proceed as follows to switch off the automatic storage function before the
adjusted storage duration has expired:
7.3
Measurement data storage
7.3.1
Displaying and editing the
measurement data storage
The contents of the manual or automatic measurement data storage can be
shown on the display and output to the interface.
Each of the measurement data storages has a function to erase the entire
contents.
Editing the data stor-
age
The storage is edited in the menu,
Storage & config
/
Data storage
. To open
the
Storage & config
menu, press the
<F1_>
[Menu] key in the measured value
display. Open the manual storage directly with the
<RCL>
key.
Open the manual or automatic storage directly with the
<RCL>
or
<RCL_>
key.
1.
Press the
<STO_>
key.
The following window appears.
2.
Using
<
><
>
, select
Yes
and confirm with
<ENTER>
.
The meter switches to the measured value display.
The automatic storage function is terminated.
Stop automatic storage?
Yes
No
Warning
01.07.2011
08:00
Back
The settings are explained here using the manual storage as an
example. The same settings and functions are available for the
automatic storage.