System 182
Relay outputs
6 - 11
ba75503e07
04/2012
Characteristic curves
Through the selection of the
Start value
and
End value
, the proportional
output can be operated with a positive or negative characteristic curve.
Positive characteristic curve:
Select the
End value
to be greater than the
Start value
.
The turn-on duration or frequency increases with an increasing
measured value (see page 6-12).
Negative characteristic curve:
Select the
End value
to be smaller than the
Start value
.
The turn-on duration or frequency decreases with an increasing
measured value (see page 6-13).
The maximum values for the pulse width or switching frequency are
assigned to the
End value
value and the minimum values for turn-on
duration or frequency are assigned to the
Start value
value.
Fig. 6-6
Positive characteristic curve
Fig. 6-7
Negative characteristic curve
Measured value
Switching frequency f or
Pulse
v
width
End value
Start value
Max.
Min.
Measured value
Start value
End value
Max.
Min.
Switching frequency f or
Pulse
v
width
Summary of Contents for DIQ/S 182-PR
Page 6: ...Contents System 182 0 4 ba75503e07 04 2012 ...
Page 10: ...Overview System 182 1 4 ba75503e07 04 2012 ...
Page 14: ...Safety instructions System 182 2 4 ba75503e07 04 2012 ...
Page 78: ...Operation System 182 4 18 ba75503e07 04 2012 ...
Page 84: ...The SETTINGS menu System 182 5 6 ba75503e07 04 2012 ...
Page 110: ...Relay outputs System 182 6 26 ba75503e07 04 2012 ...
Page 122: ...PROFIBUS connection System 182 7 12 ba75503e07 04 2012 ...
Page 124: ...Maintenance and cleaning System 182 8 2 ba75503e07 04 2012 ...
Page 130: ...What to do if System 182 9 6 ba75503e07 04 2012 ...
Page 142: ...Accessories and options System 182 11 2 ba75503e07 04 2012 ...
Page 146: ...Appendix store separately if required System 182 13 2 ba75503e07 04 2012 ...