System 182
PROFIBUS connection
7 - 9
ba75503e07
04/2012
7.5.3
Output data
(1 byte - from the PROFIBUS master to the DIQ/S 182-PR)
7.5.4
Input data
(16 bytes - from the IQ S
ENSOR
N
ET
to the PROFIBUS master)
Address
Information
Bit
Offset 0h
Sensor number (Sxx) in the IQ S
ENSOR
N
ET
(Int8)
Bit 7-0
Address
Information
Bit
Offset 0h
Sensor number (Sxx) (Int 8)
Bit 7-0
Offset 1h
Sensor status (Int8)
Bit 7-0
Offset 2h
Sensor model (Int16)
Bits 15-8
Offset 3h
Bit 7-0
Offset 4h
Status info (Int16)
Bits 15-8
Offset 5h
Bit 7-0
Offset 6h
Measuring mode (Int8)
Bit 7-0
Offset 7h
Measured value status (Int8)
Main measured value bits 7-4
Secondary measured value bits 3-0
Bit 7-0
Offset 8h
Main measured value (IEEE-754 floating point, 32-
bit)
Bits 31-24
Offset 9h
Bits 23-16
Offset Ah
Bits 15-8
Offset Bh
Bit 7-0
Offset Ch
Secondary measured value (IEEE-754 floating point,
32-bit)
Bits 31-24
Offset Dh
Bits 23-16
Offset Eh
Bits 15-8
Offset Fh
Bit 7-0
Summary of Contents for DIQ/S 182-PR
Page 6: ...Contents System 182 0 4 ba75503e07 04 2012 ...
Page 10: ...Overview System 182 1 4 ba75503e07 04 2012 ...
Page 14: ...Safety instructions System 182 2 4 ba75503e07 04 2012 ...
Page 78: ...Operation System 182 4 18 ba75503e07 04 2012 ...
Page 84: ...The SETTINGS menu System 182 5 6 ba75503e07 04 2012 ...
Page 110: ...Relay outputs System 182 6 26 ba75503e07 04 2012 ...
Page 122: ...PROFIBUS connection System 182 7 12 ba75503e07 04 2012 ...
Page 124: ...Maintenance and cleaning System 182 8 2 ba75503e07 04 2012 ...
Page 130: ...What to do if System 182 9 6 ba75503e07 04 2012 ...
Page 142: ...Accessories and options System 182 11 2 ba75503e07 04 2012 ...
Page 146: ...Appendix store separately if required System 182 13 2 ba75503e07 04 2012 ...