System 182
Relay outputs
6 - 21
ba75503e07
04/2012
6.4.8
Cleaning
Function
The
Cleaning
function enables the time controlled automatic start of the
sensor cleaning function with the aid of a relay of the Universal
Transmitter. The relay controls the DIQ/CHV valve module and
switches on or off the compressed air for the CH sensor cleaning head.
In order to set up the
Cleaning
function for a relay output, the relay
output must be linked with a sensor (see section 6.2.1).
The assigned relay always works as a closer.
The cleaning cycle consists of
Cleaning duration
and
Adjustment time
.
During the cleaning cycle the
Clean
display flashes. The outputs linked
with this sensor are frozen. The maintenance condition is active.
After the
Cleaning duration
, the relay is opened. During the following
Adjustment time
the outputs remain blocked.
The outputs linked with this sensor are only released when the cleaning
cycle is completed. The
Clean
display disappears. The maintenance
condition is finished.
Testing the
operativeness
You can test the operativeness of the cleaning system as follows:
manually open or close the relay with the
Manual control
function (see
section 6.4.9) and, while doing so, check the behavior of the cleaning
system.
Alternatively, you can test the operativeness of the cleaning system by
checking the performance of the function at the start time set up
(reference time ± interval). To perform a test immediately, you can set
the reference time so that the next cleaning cycle will start in a few
minutes (settings: see following table).
Summary of Contents for DIQ/S 182-PR
Page 6: ...Contents System 182 0 4 ba75503e07 04 2012 ...
Page 10: ...Overview System 182 1 4 ba75503e07 04 2012 ...
Page 14: ...Safety instructions System 182 2 4 ba75503e07 04 2012 ...
Page 78: ...Operation System 182 4 18 ba75503e07 04 2012 ...
Page 84: ...The SETTINGS menu System 182 5 6 ba75503e07 04 2012 ...
Page 110: ...Relay outputs System 182 6 26 ba75503e07 04 2012 ...
Page 122: ...PROFIBUS connection System 182 7 12 ba75503e07 04 2012 ...
Page 124: ...Maintenance and cleaning System 182 8 2 ba75503e07 04 2012 ...
Page 130: ...What to do if System 182 9 6 ba75503e07 04 2012 ...
Page 142: ...Accessories and options System 182 11 2 ba75503e07 04 2012 ...
Page 146: ...Appendix store separately if required System 182 13 2 ba75503e07 04 2012 ...