67
5)
Replace the built-in scanning module.
7.21
Error Code 0303
Main II Board Error Fault: Hardware 2 cannot be started; the version number appears 255
or the correct version information is not received.
7.21.1
Reasons
1)
Main II board chip is poorly soldered, resulting in failure to read normal data.
2)
There is a problem with the Main II board circuit board, causing abnormalities such as
power supply.
3)
The data connection cable of the Main II board is abnormal, resulting in the signal to
not be transmitted normally.
7.21.2 Troubleshooting
1)
Replace the cable between the Main I and Main II board.
2)
If it can't be solved, you will need to replace the Main II board.