51
2. Replace the drawer location wire and
check whether drawer functions are
restored using the debugging function.
3. Make the drawer open and close for a
switching test to check whether there is
an abnormal sound when the drawer drive
motor is working. If there is any
abnormal sound, it may be caused by
interference of components or damage to
the drawer drive motor. It is necessary to
adjust or replace the drawer drive motor.
Take out the whole mechanical module.
4. Disassemble the four screws on the
opposite side and remove the assembly
parts.
5. Unscrew the screw with an inner hexagon
wrench, remove the bronze gear.
Summary of Contents for InSight V-CHEM
Page 1: ...Veterinary Chemistry Analyser Service Manual...
Page 9: ...3 Back View Right View...
Page 42: ...36...