57
3)
Replace the Main II board.
1. Unplug the connection line between the
Main I and Main II board. Screw down
the four screws as shown in the figure.
Remove the Main I board. (If there is hot
melt glue at the connector, it can be
removed easier by applying alcohol.)
2. Remove all the connection wires on the
Main II board. Use a wrench to
disassemble the four columns as shown in
the figure.