46
13. Unscrew two screws as shown in the
picture and remove the optical
assembly downward.
14. Replace with a new whole optical module
which is supplied by Woodley Equipment
Company.
Notice:
The whole optical module is
tested for-all performance parameters
by Woodley Equipment Company.
7.6
Error Code 0108
Location LED Abnormal: Change the channel 1 for IR acquisition and the collected value
is compared with the default IR threshold. When the attenuation is more than 1/3, an error
is reported.
7.6.1
Reasons
•
LED light intensity attenuation
•
LED was damaged
•
LED power supply is abnormal
7.6.2
Troubleshooting
Please follow the steps in order. After each step is finished, retest whether the fault is cleared
or not. If not, proceed to the next step.