13
6.
The first function of the
Debug
button
is
the
temperature
calibration
function.
7.
Touch the
Next Page button
to check
the basic functions of the drawer
drive motor and centrifugal motor.
Press
Open
, put in a new disc or any
tested disc, then press
Close
. Wait
until the drawer is completely closed
then press
Motor Start-up
.
The
number shows on the
Motor Speed
place is the current speed. Once the
motor runs stably at the speed of
1200±50rpm, the motor speed is
normal.
Note:
You can press
Motor Stop
to stop execution.