22
4 Mounting
WIKA Operating Instructions - Differential pressure transmitter DPT-10
. Mount DPT10 directly to the vessel
. Minus side is open to the atmospheric pressure
+
min.
p
atm
p
atm
1
2
+
–
Fig. 20: Measurement setup, level measurement in the open vessel
1 DPT10
2 Minus side is open to the atmospheric pressure
. Mount DPT10 below the lower measurement connection so that
the effective pressure lines are always filled with liquid
. Connect minus side always above the max. level
. For measurements in products with solid content such as e.g.
dirty liquids, the installation of separators and drain valves is
recommended to enable collection and removal of debris and
sediment.
+
–
min.
max.
1
3
4
3
4
5
1
2
Fig. 21: Measurement setup, level measurement in closed vessel
1 Blocking valves
2 DPT10
3 Precipitator
4 Drain valves
5 3-fold valve block
In open vessels with
single chemical seal
In closed vessels with ef-
fective pressure lines