6 Commissioning, operation
74 WIKA Operating Instruction, Pascal ET, ET-P, ET/IS, ET-P/IS
GB
“Dry well Baud rate”: set the Baud rate value that must be the same used by the dry well calibrator.
Note
: some dry well calibrators require the interface converter kit.
“Stab. Wait. Time”: set the stability waiting time in the hh:mm:ss format. The Pascal calibrator, after
setting the temperature point for the report for the serial interface, will wait this dead time before
starting to detect the temperature stability value.
“Stab. Time” and “Stab. Band”: the algorithm for detecting the stability value will consider the
temperature stable by using the two set parameters. When the instrument will detects a stable
temperature, the report will store the values of the thermometers and then will proceed to the next
set point.
The “Stab. Band” parameter indicates the temperature range where the temperature value must
not fall within for the set time. This allows to detect a stable temperature value.
The value describes above depend of the type of the connected dry well calibrator and its setting
specifications.
Consider the following chart: