5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System
Rx Only – Z370147 Rev. C 1008
5-26
Table 5.8 – PHACO – AMO Default Settings
Parameter
Available
Settings
AMO Default Settings
Phaco 1
Phaco 2
Phaco 3
Phaco 4
Bottle Height
0–104cm
0–42 inches
66 cm
26 inches
76 cm
30 inches
76 cm
30 inches
76 cm
30 inches
Continuous
Irrigation
On, Off
Off
Off
Off
Off
Pump Selection
Venturi
Peristaltic
Peristaltic
Peristaltic
Peristaltic
Peristaltic
Max Vacuum
Peristaltic
0–650 mmHg
75 mmHg
300 mmHg
400 mmHg
350 mmHg
Max Vacuum
Venturi
0-600 mmHg
75 mmHg
300 mmHg
400 mmHg
350 mmHg
Venturi Pump
On, Off
Off
Off
Off
Off
Passive Reflux
Yes, No
Fusion™
Fluidics Pack
Yes
Yes
Yes
Yes
Fusion™ Dual
Pump Pack
Yes
Yes
Yes
Yes
Initial Mode/
Submode
Any of the
Submode
Names
Phaco 1
Phaco 1
Phaco 1
Phaco 1
Max Flow/
Aspiration Rate,
Unoccluded
(Peristaltic)
10–60 cc/min
18 cc/min
24 cc/min
18 cc/min
24 cc/min
Fluidics Mode
Panel, Linear
Panel
Aspiration,
Linear
Vacuum
Panel
Aspiration,
Panel
Vacuum
Panel
Aspiration,
Linear
Vacuum
Panel
Aspiration,
Linear
Vacuum
Occlusion
Threshold
0–Max Vac.
25 mmHg
140 mmHg
60 mmHg
80 mmHg
OCCLUSION
MODE
TM
Phaco
On, Off
Off
On
On
Off
CASE Mode
On, Off
Off
Off
On
On
CASE Upper
Threshold.
0–Max Vac.
100 mmHg
250 mmHg
325 mmHg
300 mmHg
CASE Lower
Threshold.
0–Max Vac.
70 mmHg
175 mmHg
200 mmHg
200 mmHg
Up Time
0–2000 ms
300 ms
300 ms
300 ms
300 ms
CASE Vac
120–650
mmHg
85 mmHg
225 mmHg
250 mmHg
250 mmHg
CASE
One-Touch
-2, -1, STD,
+1, +2
STD
STD
STD
STD
Peristaltic Pump
Ramp
10–100%
50%
60%
50%
70%
Power Delivery
Mode
(Unoccluded and
Occluded)
Continuous
Short Pulse
Long Pulse
Low Power
Pulse
High Power
Pulse
Single Burst
Multi-Burst
Continuous
Burst
Continuous
Continuous
Continuous
Continuous
Short Pulse Rate
1–14 pps
6 pps
6 pps
6 pps
6 pps
Long Pulse Rate
1–6 pps
4 pps
4 pps
4 pps
4 pps