5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System
Rx Only – Z370147 Rev. C 1008
5-8
Using PHACO
1. Press
PHACO
.
2. Press
Phaco 1
,
Phaco 2
,
Phaco 3
, or
Phaco 4
to select the submode. The main
panel shows all of the settings for Aspiration Rate, Vacuum and Power. Each
submode has different default settings.
3. Press the footpedal to activate phacoemulsification. The Aspiration Rate,
Vacuum and Power control panels indicate the associated levels throughout the
procedure. The Footpedal icon in the upper right corner of the screen shows the
position of the footpedal.
How to Program PHACO
1. Press the
PHACO
button.
2. Press the desired submode button.
Phaco 1
is selected by default.
Note: You can set your Initial Mode/Submode on the Phaco Configuration screen.
The Initial Mode/Submode is then your default setting. (Refer to Program
Configuration, in this chapter.)
3. In the control panels in the main panel, click the
Up
or
Down
arrows to
increase or decrease Aspiration Rate, Vacuum, or Power.
Note: If the Venturi pump is On, only the Vacuum and Power Settings are shown.
CASE is not accessible when the Venturi pump is On.
4. Repeat for each submode you want to program.
5. If OCCLUSION MODE™ Phaco or CASE is desired, press
in the
left column to activate the FUSION™ Mode screen. When OCCLUSION
MODE™ Phaco is active, the
On
button is green.
Figure 5.7 – Fusion™ Fluidics Screen