5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System
Rx Only – Z370147 Rev. C 1008
5-35
End Case
Purge
– Select
Purge
to remove all of the fluid from the irrigation and aspiration
tubing before you remove the tubing cassette.
Next Case
– Select
Next Case
to install a new tubing cassette.
Note: Use a new bottle of balanced salt solution at the start of each case.
Figure 5.27 – End Case Screen
The timers in the upper left corner indicate Effective Phaco Time (EPT), Ultrasonic
(U/S) Time in Footpedal Position 3, and Ellips EPT if Ellips™ Technology is On.
Effective Phaco Time is ultrasound time as a weighted total that takes into account
the amount of power being used:
• at 100% power: 1 sec. U/S Time = 1 sec. EPT
• at 50% power: 1 sec. U/S Time = .5 sec. EPT
Current Case
Review Case
– shows the graphical history of the surgery. (Refer to Figure 5.28
Review Case Screen.)
1. Select a shorter time interval to adjust the Time Line Duration and to see
greater case details. For example, 10 Sec shows more detail than 15 Min.
2. Select a longer time interval for an overview of the surgery or sessions, such as
1 hour or 2 hours. This view provides less detail, but gives you a better
indication of the trends over a period of time. Use the
<<
or
>>
buttons to
change the time intervals. You cannot use the arrow buttons when you view an
active case.