5 • Anterior Segment Surgery Operating Modes
WHITESTAR SIGNATURE™ System
Rx Only – Z370147 Rev. C 1008
5-5
The irrigation tubing runs through the tubing cassette and irrigation is controlled by
the footpedal. When the footpedal is pressed, the pinch valve is opened and the
irrigation fluid flows. Irrigation runs in footpedal Positions 1, 2 and 3.
To program IA submodes:
1. On the left panel, press
IA
.
2. Select the desired IA submode.
IA 1
is selected by default.
3. Press the
Up
or
Down
arrows in the control panels on the main panel to
increase or decrease Aspiration Rate or Vacuum.
Note: If the Venturi pump is
On
, only the Vacuum Settings are shown. To turn
the Venturi pump setting on, use the
IA Submode Configuration
screen.
4. Press the
Settings
button in the control panels on the main panel to change
Aspiration Mode or Vacuum Mode,
5. Press
Finished
to close the
Settings
window.
6. After all the submodes are programmed, select another mode.
7. If no other changes are needed press
End Case
to access the
SAVE
buttons and
save the settings.
Figure 5.4 – Irrigation/Aspiration Mode