Instruction for use Radiant Warmer CERAMOTHERM 1000
D2027_02
7/25
First switch-on
First of all make sure that the radiant warmer is connected to the power supply. The pilot lamp for the safety
cut-off at the bottom side of the device indicates the readiness for operation.
Select an intensity of 10 mW/cm² with knob (13). The selection is indicated as LED-spot in display (12).
The green pilot lamp for the heating power (11) flashes according to the intensity selection, i.e.:
•
Low radiation intensity
= short flashes
•
High radiation intensity = long flashes.
Depending on the ambient temperature the ceramic heating elements require approx. 4 to 6 minutes until the
selected intensity of 10 mW/cm² is reached.
11
6
2
1
5
Switch-on the light with button (1). The green pilot lamp (5) and the
illumination of the patient area light.
Switch-on the heating with button (2). The green pilot lamp (6) lights.
A self-test is performed. A tone sounds and all displays light-up for
2 seconds.
Check during the 2 seconds whether all displays light up and the
tone sounds.
When the self-test is finished successfully, the radiant warmer starts
heating-up. The green pilot lamp for heating power (11) lights.