Instruction for use Radiant Warmer CERAMOTHERM 1000
D2027_02
14/25
Double-joint wall arm
Double-joint ceiling arm
Height-adjustable mounts
Application to babies and defenseless persons under surveillance only!
The safety distance of 650 mm between the bottom line of the radiant warmer and the patient pad must be
maintained. If necessary it must be secured in a mechanical way, i.e. by a stop in the height-adjustable
spring-balanced arm.
Double-joint arm
Adjustable
distance to the patient pad.
Horizontal swiveling 180° (wall arm) respectively 360° (ceiling arm), adjustable stop for highest position,
rotation of warmer 360°, vertical adjustment ± 45°.
Rotate respectively swivel the warmer to the desired position. Use the two handles and do not touch the
safety guard. Take care that wall or ceiling arm extensions will not clash with other devices in the ambience,
as e.g. ceiling lights, mobile stands, etc.
When the radiation shall be focussed to the patient in diagonal instead of vertical direction, unfasten the
release handle (
H
) by anti-clockwise turning and adjust the warmer to the desired direction. Afterwards
fasten the release handle again by clockwise turning.
H
H