Instruction for use Radiant Warmer CERAMOTHERM 1000
D2027_02
23/25
Height-adjustable mounts
Application to babies and defenseless persons under surveillance only!
The safety distance of 650 mm between the bottom line of the radiant warmer and the patient pad must be
maintained. If necessary it must be secured in a mechanical way, i.e. by a stop in the height-adjustable
spring-balanced arm.
Double-joint arm
Adjustable
distance to the patient pad.
Horizontal swiveling 180° (wall
arm) respectively 360° (ceiling arm), adjustable stop for highest position,
rotation of warmer 360°, vertical adjustment ± 45°.
Operating distance to patient pad: 800 mm
Safety distance to the patient pad: 650 mm
CERAMOTHERM 1100
CERAMOTHERM 1200
Double-joint-
wall arm
R1 = 750 mm, R2 = 800 mm
Order No. WY1107
Order No. WY1207
Special length R1 = 950 mm
Order No. WY1847
Order No. WY1847
Double-joint
ceiling arm
R1 = 750 mm, R2 = 800 mm
Distance between patient pad and bottom line of
solid ceiling 1800 to 2600 mm
Order No. WY1108
Order No. WY1208
Double-joint
ceiling arm
R1 = 750 mm, R2 = 800 mm
Distance between patient pad and bottom line of
solid ceiling above 2600 mm
Order No. WY1118
Order No. WY1218
Special length R1 = 950 mm
Order No. WY1848
Order No. WY1848
Weight with wall arm respectively ceiling mount /
torque:
Double-joint-
wall arm
18.7 kg / 197 Nm
21.2 kg / 236 Nm
Double-joint-
ceiling arm
Distance between patient pad and bottom line of
solid ceiling 1800 to 2600 mm
28.4 kg / 186 Nm
30.9 kg / 223 Nm
Double-joint-
ceiling arm
Distance between patient pad and bottom line of
solid ceiling above 2600 mm
38.0 kg / 186 Nm
40.0 kg / 223 Nm
Intensity selection /
irradiated area at operating distance
2 to 18 mW/cm² / 480 x 580 mm
2 to 22 mW/cm² / 480 x 730 mm
Double-joint wall arm,
Extension arm R1 = 750 mm, height-
adjustable spring balanced arm R2 = 800 mm.
Double-joint ceiling arm,
Extension arm R1 = 750 mm, height-adjustable
spring balanced arm R2 = 800 mm.
Mounting at the solid ceiling.