Instruction for use Radiant Warmer CERAMOTHERM 1000
D2027_02
20/25
CERAMOTHERM 1100
CERAMOTHERM 1200
Max. power input
690 W / 3 A
900 W / 3,9 A
Heating element(s), ceramic
1 x 600 W
2 x 400 W each
Intensity selection / irradiated area
at distance to patient pad:
650 mm
2-28 mW/cm² / 390 x 520 mm
2-30 mW/cm² / 390 x 680 mm
700 mm
2-24 mW/cm² / 430 x 540 mm
2-30 mW/cm² / 430 x 700 mm
750 mm
2-20 mW/cm² / 460 x 560 mm
2-26 mW/cm² / 460 x 720 mm
800 mm
2-18 mW/cm² / 480 x 580 mm
2-22 mW/cm² / 480 x 730 mm
850 mm
2-16 mW/cm² / 500 x 600 mm
2-20 mW/cm² / 500 x 750 mm
900 mm
2-14 mW/cm² / 520 x 620 mm
2-18 mW/cm² / 520 x 770 mm
Illumination
2 fluorescent tubes 11 Watt each
2 fluorescent tubes 18 Watt each
Classification
Protection class
1
Device type
B
MDD class
IIa
Standards
The device complies with
EN 60601(edition 1996)
EN 60601-2-21, 1995-12 resp. modification A1 1998-0
Certification