Instruction for use Radiant Warmer CERAMOTHERM 1000
D2027_02
17/25
Cleaning and disinfection
The radiant warmer should be cleaned and disinfected for every new patient, however, at least once per
week.
Cleaning agents and disinfectants
Never use inflammable disinfectants!
Do not use scouring cleaning agents. Use only disinfectants on the basis of aldehydes and quaternary
ammonium merges. Unsuitable are tincture of iodine 5 %, carbolic acid, spirit, ether, acetone and other
alcoholic agents as well as disinfectants on the basis of halogen-splitting merges, strong organic acids and
oxygen-splitting merges.
When choosing a substance generally the manufacturer's recommendations must be observed. The
manufacturer of the substance is hold liable for the application field and possible material damages.
The following disinfectants can be recommended:
Incidur®
Henkel Hygiene, Düsseldorf / Germany
Gevisol
Schülke & Mayr, Norderstedt / Germany
Buraton 10F
Schülke & Mayr, Norderstedt / Germany
Bacillotox
Bode, Hamburg / Germany
Cleaning and disinfecting
After use switch-off the radiant warmer and wait until it has cooled-off. Disconnect the power plug.
Clean the device with a cloth slightly saturated with cleaning agent and wipe it dry.
No liquid must enter into the top of the device housing.
The ceramic heating elements must not be treated with cleaning agents or disinfectants!
After disinfection let the radiant warmer dry for at least 1 hour. Do
not
switch it on in order to accelerate the
drying process.
After cleaning and disinfection check the functions.
Waste disposal
Fluorescent tubes:
Hazardous waste.
Batteries:
Hazardous waste.