Instruction for use Radiant Warmer CERAMOTHERM 1000
D2027_02
15/25
H
1
2
Mobile radiant warmers
Application to babies and defenseless persons under surveillance only!
Height-adjustable mobile stand,
4 anti-static castors, 2 with kickstop,
rotation of warmer 360°, vertical adjustment ± 45°.
Unfasten the knob of the height-adjustment (1) by anti-
clockwise turning and adjust the stand to the lowest
position. Fasten the knob again by clockwise turning.
Release the kickstop of the castors, move the device to
the desired position and brake the castors again. The
device must be moved in lowest position only.
The radiant warmer is calibrated to a distance of
650 mm (safety distance)
to the patient pad. Now
adjust the height in a way that the
red distance
mark (2
) is at the level of the patient pad. Take care that
the knob of the height-adjustment (1) is fastened again
by clockwise turning.
Rotate respectively swivel the warmer to the desired
position. Use the two handles and do not touch the
safety guard. When the radiation shall be focussed to
the patient in diagonal instead of vertical direction,
unfasten the release handle
(
H
) by anti-clockwise
turning and adjust the warmer to the desired direction.
Afterwards fasten the release handle again by clockwise
turning.