Mounting / Installation
Depending on the supplied model the radiant warmer CERAMOTHERM 1000 can be mounted at the wall or
ceiling by qualified persons by observing our instructions for installation. The radiant warmer can also be
mounted at a mobile stand. Please refer to the remarks for the individual mounts on pages 12 to 15 and 21
to 24.
For installation it has to be considered that heating or air conditioning devices can change the effect of the
radiant warmer. Therefore a distance of 1 m should be kept clear. This distance should also be kept between
two radiant warmers. Furthermore the radiant warmer should not be placed in draughts, at cold walls or
windows as well in direct sunlight.
In case of radiant warmers with flexible arrangement to the patient pad take care that no items are placed
within the adjusting/moving radius of the radiant warmer.
No fluids must be placed above the radiant warmer which in the case of fault might flow into the device.
For fire prevention a distance of min. 50 cm must be kept clear between the radiant warmer and any
inflammable material (e.g. curtains), under consideration of the moving radius of the radiant warmer in case
of an adjustable mount.
The radiant warmer provides an electronical control which allows to exactly select the radiation intensity
emitted to the patient pad. The selection is decisively dependent from the distance between the radiant
warmer and the patient pad. Therefore the mounting height must be determined with corresponding care.
The factory distance calibration is obvious from the sticker which is placed between the rear faceplate and
the device housing. If the radiant warmer shall be mounted at a different distance, this should only be done
with the consent of the manufacturer or authorized supplier because in that case the device must be
recalibrated to the new distance.
For ergonomical reasons different standard respectively operating distances apply for the various mounts.
The corresponding distance as well as the weights and torques can be taken from the charts below the
individual mounts, pages 21 to 24.
A safety distance of 650 mm between the bottom line of the radiant warmer and the patient pad must be
maintained. Height-adjustable wall and ceiling mounted radiant warmers must be mounted in a way that at
highest patient level and lowest position of the radiant warmer a safety distance of 650 mm is maintained.
When calculating the mounting height the standard respectively operating distance plus the thickness of the
pad must be considered in order to verify the correct distance to the patient.
A direct installation to the electrical mains must only be done by a qualified electrician.