Table 3: inFlow™ ACE Troubleshooting (Continued)
Issues
Possible Causes
Solutions
The sample container is filling too
quickly.
The inFlow™ ACE is not collecting the
correct sample volume.
The inFlow™ ACE may be set at a faster
sampling frequency than desired.
The relief on the inFlow™ ACE may be set
too low.
Welker recommends the Welker
Checkpoint™ Sample Bite Verification
Panel for verifying the sample volume of
the inFlow™ ACE. See
for instructions on
adjusting the sample volume.
Adjust the inFlow™ ACE to sample at the
desired rate. Ensure that the calculations
used to determine the sample frequency
are correct (
Check the setting on the relief and adjust
as necessary.