32
IOM-161
MODEL: INFLOW™ ACE WITH AI CONTROL™
REV: D
13839 West Bellfort Street, Sugar Land, TX 77498
welker.com
Service Department 281.491.2331
Reassembling the Upper Housing
42.
Carefully slide the external adjustment body onto the insertion shaft. The lock collar grooves in the insertion shaft should
be accessible on both sides of the external adjustment body.
43.
Install the lower lock collar to the groove below the external adjustment body and align the pin hole in the lock collar
with the pin in the external adjustment body.
The lower lock collar must be installed to the insertion shaft assembly before the upper lock collar.
44.
Install the upper lock collar to the groove above the external adjustment body and align the pin hole in the lock collar
with the pin in the external adjustment body.
45.
Tighten the setscrews in the upper and lower lock collars to secure the adjustment assembly to the insertion shaft.
46.
Carefully insert the insertion shaft assembly through the packing gland body, taking care not to scratch the shaft.
47.
Slide the gland cover onto the packing gland body, and then return the plug to the lubrication port.
48.
Slide the dustcover onto the packing gland body.
49.
Carefully slide the motor housing onto the insertion shaft, taking care not to scratch the shaft.
50.
Carefully screw the motor housing into the packing gland body.
51.
Return the top cap to the top of the insertion shaft and carefully slide the top cap down to the motor housing.
52.
Bolt the top cap in place.
53.
Set the upper housing and insertion shaft assembly aside.