7. Refit the grommet
8. Refit the board (see “6.7 Changing the board”
on page 36, steps
9. Close the device (see “6.4 Close the device”
10. Perform a final check (see "3. Final check" on
6.13 Replace lower part of housing
Materials and tools required
• ESD-protected workplace
• Size 2 Phillips screwdriver
• Open-ended wrench, size 10,
• Open-ended wrench, size 13,
• Open-ended wrench, size 14,
• Open-ended wrench, size 22,
• Torque wrench 50 ± 5 Ncm,
• Torque wrench 200 ± 10 Ncm,
• Special counter tool G 3/8 (WM 22827) and special wrench, size 17 (WM 22828) from special
tool set WM 15349,
• Allen key for grooved nut M1 x 0.5 (WM 28166)
• Vise with protective jaws
• Flat-nose pliers
1. Open the device (see "6.3 Open the device"
2. Remove the pneumatic block (“6.8 Replacing
the pneumatic block” on page 37, steps
to
3. Disconnect cable harness
communication socket
4. Remove the nut from communication socket
rear of the device.
5. Undo the nut on the MEDUtrigger socket
from the outside using the Allen key for
grooved nuts.
6. Remove the MEDUtrigger socket
from the