Repair information and instructions
41
6.11 Changing upper part of housing
Tools and equipment required:
• ESD-protected workplace,
• Phillips screwdriver, size 2,
• Tubular hexagon box spanner 10 mm,
• Torque spanner 50 ±5 Ncm,
• Torque spanner 200 ±10 Ncm,
• Special tool WM 22829 from special tool set WM 15349,
• Flat-nose pliers.
1. Open the device (see “6.3 Open the device”
2. Remove the board (see “6.7 Changing the
board” on page 36, steps
You have now removed all the parts. You can start
reassembling.
4. Insert the grommet
5. Refit the board (see “6.7 Changing the board”
on page 36, steps
6. Close the device (see “6.4 Close the device”
7. Perform a final check (see "3. Final check" on