44
Repair information and instructions
7. Use a sharp object to push the blanking plug
of the interface connection out of lower part
8. Take O-ring
out of lower part of the housing
9. Turn the special counter tool onto pressurized
gas connection
10. Clamp the special counter tool in a vise.
11. Turn the nut of the special counter tool against
the pressurized gas connection using an open-
ended wrench (size 22).
12. Unscrew elbow connector
using an open-
ended wrench (size 13).
13. Use the special wrench (size 17) to slacken nut
14. Take metal piece
off upwards.
15. Put a new lower part of housing
on
pressurized gas connection
16. Push metal piece
housing onto the connection.
17. Screw nut
(size 17) on the inside of the
connection tight.
18. Fix elbow connector
on the connection.
19. Undo the nut of the special counter tool using
the open-ended wrench (size 22).
20. Open the vise.
21. Turn the special counter tool off pressurized
gas connection
22. Pass the cable for the MEDUtrigger socket
into the inside of the housing.
23. Tighten the nut on the MEDUtrigger socket
from the outside using the Allen key for
grooved nuts.
24. Pass the cable for communication socket
into the inside of the housing.
25. Tighten the nut on communication socket
26. Connect the cable for communication socket
27. Push blanking plug
into the interface
connection from the outside.
28. Put O-ring