26
Troubleshooting
5. Troubleshooting
Fault
Cause
Remedy
MEDUMAT Easy CPR does
not start up when switched on
A battery is exhausted
(3.8.1, Page 10) and replace if
necessary (4.2, Page 16)
Fuse defective
Ribbon cable
X1000
of fascia film is
faulty or not connected
Check plug-in connection and cable
(6.7, Page 26)
If necessary, replace upper part of
housing (6.11, Page 31)
defective
Change fascia film (6.12, Page 32)
Board defective
Magnetic valve defective
Change magnetic valve (6.9,
Page 29)
MEDUMAT Easy CPR will not
switch off
Operating error
Keep On/Off button
at least 2 seconds
defective
Change fascia film (6.12, Page 32)
MEDUMAT Easy CPR is
functioning, but without any
displays
Pressure gauge hose on
MEDUMAT Easy CPR or on patient
valve has slipped off
Check pressure gauge hose
Kink in pressure gauge hose
Pressure gauge hose within device is
kinked or has slipped off
V
t
too high
Measured without 10 mbar back
pressure
Set to 10 mbar back pressure
V
t
not correct
Measuring device not calibrated
Calibrate measuring device
Input pressure > 6 bar
Reduce system setting to less than
6 bar
Patient valve not in order
Check membranes and O-ring,
replace if necessary (Chapter 6.7 of
operating instructions)
Potentiometer wrongly adjusted
Readjust potentiometer (6.10,
Page 30)
Adjuster knob out of adjustment
Readjust adjuster knob (6.10,
Page 30)
Pneumatic block leaking