Repair information and instructions
35
6.6 Changing the speaker
Tools and equipment required
• ESD-protected workplace,
• Phillips screwdriver, size 1,
• Tubular hexagon box spanner 10 mm,
• Torque spanner 50 ±5 Ncm,
• Torque spanner 200 ±10 Ncm,
• Special tool WM 22829 from special tool set WM 15349.
1. Open the device (see “6.3 Open the device” on page 30, steps
the housing
and disconnect the speaker ca-
ble plugs.
3. Take the new speaker
and connect the
speaker cable plugs.
part of the housing
speaker must be located in the corresponding
groove in the lower part of the housing.
5. Close the device (see “6.4 Close the device”
on page 32, steps
).
6. Perform the final check (see “3. Final check” on
Demand-
flow
Centering
lug