Troubleshooting
No Stenosis alarm when
patient valve closed during
functional check (see
“Checking the breath volume”
in the operating instructions)
Patient valve not working properly
Check lip membrane
The MEDUtrigger does not
trigger a respiratory stroke
(CPR mode only)
Respiratory stroke triggered during
the inspiration and expiration phase
Wait until the expiration phase is
over to trigger the respiratory stroke
(corresponds to the length of the
inspiration phase)
The MEDUtrigger is not
working
MEDUtrigger not correctly
connected
Check whether the MEDUtrigger
connector is correctly positioned and
correct it
Fault
Cause
Remedy