38
Repair information and instructions
9.
Press the catch towards the elbow connector
and pull the pneumatic block
off the
tube
.
When you remove the pneumatic block from
the lower part of the housing, the O-ring
will fall out of the connection. Make sure that
does not fall out of the relief out
-
let valve.
10.
When inserting the new or replacement pneu
-
matic block
, insert the tube
in the elbow
connector.
In the process, ensure that the sole
-
noid valve control cable and cable harness
are routed under the pneumatic block through
to the right-hand side.
11.
Turn the lower part of the housing over, hold it
and the pneumatic block tight and screw up
both screws
12.
Lubricate the O-rings of the elbow connector
2
lubricant (WM
14298).
Make sure that the compensating hole remains
free
.
13.
from the pneumatic block
.
14.
Insert the elbow connector
in the pneumatic
15.
in the pneumatic block
.
16.
Attach cable for the
MEDUtrigger
pneumatic block
.
17.
Plug in the potentiometer cable.
18.
Plug in th
e magnetic valve control cable.
19.
Close the device
(see “6.4
Close the device
”
on page
32)
.
(see “3.
Final check
” on
page
8)
.
Catch
Magnetic valve control cable
Compensating
hole
Magnetic valve
control cable
Potentiometer
cable