44
English
7. Malfunctions and
Correction
Malfunction
Cause
Correction
Operation indi-
cator of ther-
apy unit not
illuminate.
No supply volt-
age available.
Check power supply
cable is securely
connected. Possibly
check function of
outlet by connecting
another piece of
equipment (e.g.
lamp).
Breath humidi-
fier is not
warming up.
Breath humidifier
is not activated.
Switch on breath
humidifier, see 4.2,
page 20.
Defect in electron-
ics.
Send the breath
humidifier with the
therapy unit to your
specialist dealer for
repair.
The breath
humidifier is
leaking.
The seal of the
heating element
Replace the seal,
see 7.1, page 46.
The seal of the
housing base
is
defective.
Replace the seal,
see 7.2, page 48.
Cracks in base.
Replace the base.
No or incor-
rect pressure
produced at
mask.
Water in pressure
measurement
opening.
Dry pressure measure-
ment opening,
see 5., page 30.
Pressure measure-
ment tube
correctly fitted to
top section.
Fit the pressure
measurement tube to
the appropriate con-
nection stubs, see
5.2, item 6..
No supply volt-
age available.
Check power supply
cable is securely
connected. Possibly
check function of
outlet by connecting
another piece of
equipment (e.g.
lamp).