3. Wipe out the housing parts with a clean
cloth and rinse well several times with
water.
4. Then dry all parts carefully.
5.4 Disinfection
The breath humidifier can be disinfected.
Please observe the operating instructions for
the disinfectant used. You are recommended
to wear suitable gloves (e.g. household or dis-
posable gloves) during disinfection proce-
dures.
•
The procedure for disassembly and as-
sembly of the breath humidifier is as for
cleaning.
•
After being disinfected, the breath humid-
ifier has to be thoroughly rinsed with dis-
tilled water and dried.
•
The outside of the breath humidifier is
cleaned by simply wiping down with dis-
infectant.
We recommend replacing plastic parts after
max 2 years' service for reasons of hygiene.
5.5 Change in patients
If the device is to be subjected to a hygiene
treatment for another patient, perform the
steps described below:
•
Disinfect all parts.
•
If plastic parts and the heating element
are dirty or covered in limescale, replace
them.