Repair information and repair instructions
Fitting lid
1. Place the ends of the springs
holes of a new device lid.
2. Slide the pins
from the side far enough into the
3. Replace the lid
, so that the other end of the
springs are situated between the encasing halves.
5. Close the device lid.
8.6 Close the device
1. Hold the lower encasing part
2. Plug the blue cable into the printed circuit board
(Labelled
HDQ
).
3. Plug the black cable into the printed circuit board
(Labelled
ACCU -
).
4. Take the pad plug cable
into the printed circuit board.
5. Press the capacitor into the compartment of the
lower encasing part. Guide the cable in through
the slit.
6. Plug the cables for capacitor
onto the board.
Ensure that po/- is correct (see labeling on the
board and on the capacitor)! If the cables are the
wrong way round, the top and bottom parts of the
device will not fit together properly.
7. Plug the red cable into the printed circuit board
(Labelled
ACCU +
).
8. Plug the speaker cable into the printed circuit
board.
9. Place the lower encasing part
Please make sure that the cables do not get clamped.
10. Now screw the upper encasing part tightly with
the ten screws
11. Set the sealing plugs
in place.
12. Turn the device around the right way again and
fit the lid (see "8.5 Exchange the device lids" on
page 27).
13. Check the device (see "4. Checking the device"