40
Technical data
Power supply
*
For new Rechargeable Battery or Battery-Pack, 20 C.
**
At Low-Energy settings
*** At the lowest volume
Version
Battery-Pack
Rechargeable Battery-Pack (optional)
Type
LiSO
2
Li-Ion
Dimensions LxBxH (in mm)
148.6 x 71.6 x 32.6
Weight:
400 g
Shock capacity*,**:
up to 200 shocks
up to 100 shocks
Minimum capacity
100 shocks
–
Monitoring capacity*,***:
up to 18 hours
up to 9 hours
Rated capacitance:
3800 mAh
Rated voltage:
11.2 V
12.4 V
Fuse
–
16 A
Stand-by-Time*:
Minimum Stand-by-Time:
up to 5 years
4 years
Maximum charging time
–
< 4 h
Maximum charging power
–
1,2 A
Transport/Storage:
Temperature range:
Air humidity:
–
- 40˚C to +85˚C
0% to 90%
Service life
–
> 300 full charges
Battery charger (optional)
Dimensions LxBxH (in mm)
108 x 65 x 77
Mains voltage
100-240 ~V
Mains frequency
50/60 Hz
Output voltage
15 V
Weight
230g
Ambient temperature
Operation
Transport/Storage
0˚C to 40˚C
-20˚C to +85˚C
Fuse
T2A 250 V
Short circuit proof
Durable
Protective function
Voltage surge protection
Charger plug
Quadripolar
Secured against incorrect polarity
Mains plug
Exchangeable, for different countries
Electromagnetic compatibility
see “Dimensions/Environment/Norms”