7. Malfunctions and Rectification
Malfunctions
Cause
Rectification
In stand by mode (lid closed) a
signal tone is set off every four
minutes.
A malfunction has been
established while the device
underwent the monthly/daily self
tests.
Check which LED blinks and the
procedure, according to the
respective LED described in the
following table.
When the lid is open the device
can not turn on.
Battery capacity is exhausted.
Insert new Battery/Rechargeable
Battery Pack (see chapter „Energy
Supply“ in the operating
instructions MEDUCORE Easy).
Damaged connection between the
energy supply contacts and main
circuit board
Opening device and checking
connections (Chapter 8.3,
Page 25).
Main circuit board defective.
Open up device and replace the
main circuit board (Chapter 8.9,
Page 30).
Upon opening the lid the
following message can be heard:
“Device is not ready for use”
Device messages with EasyView
voice prompting. If EasyView
“Malfunction in the main circuit
board” is displayed, then replace
the main circuit board
(Chapter 8.9, Page 30).
Upon opening the lid the
following message can be heard:
“Battery is low”
Battery power levels are low.
Finish off current treatment and
then replace batteries (see
Chapter “Energy Supply” in the
operating instructions
MEDUCORE Easy).
The red LED will light up (flashing
when on stand by and constant
when in operation).
If the battery power levels are low
then fewer than 3 shocks can be
delivered.
Replace battery
(see Chapter
“Energy Supply” in the operating
instructions of the
MEDUCORE Easy), to enable
continued treatment.
Device is not ready for operation,
self test has detected a
malfunction.
Finish off current treatment. Then
read out malfunction protocol via
the EasyView. If EasyView
“Malfunction in the main circuit
board” is displayed, then
exchange main circuit board
(Chapter 8.9, Page 30).
The green and yellow LEDs light
up (flashing when on stand by
and constant when in operation).
The battery/rechargeable battery
power levels are low and barely
10/6 shocks can still be
delivered.
Finish off current treatment and
then replace batteries (see
Chapter “Energy Supply” in the
operating instructions
MEDUCORE Easy).
The yellow LEDs light up (flashing
when on stand by and constant
when in operation).
A less critical malfunction has
been detected, e.g. time is
incorrect.
Finish off current treatment. Then
read out malfunction protocol via
the EasyView and if necessary
correct malfunction e.g. reset
clock.