Repair information and repair instructions
25
Danger of death! Danger of electric shock! If no discharg-
ing unit is available, work on the MEDUCORE Easy must
be carried out at a protected high-voltage workstation.
Before carrying out any repair, it is essential to discharge
the capacitor as described in Section “8.3 Discharge high
voltage capacitor and open up device” on page 25, oth-
erwise fatal injuries may be sustained.
8.3 Discharge high voltage capacitor and open up device
Preparation
Danger! Risk of electric shock!
If a malfunction occurs in the device during a functional
check, then it may not be used.
1. Remove the Battery-Pack/Rechargeable Battery
Pack from the battery compartment
MEDUCORE Easy.
2. Lie the MEDUCORE Easy with the red lid
facing
down
on a non-slip surface.
3. Remove the sealing plugs
4. Loosen and remove the ten screws
5. Take the MEDUCORE Easy in both hands and
hold both encasing elements together. Lie the
MEDUCORE Easy with the red lid
facing upward
in the discharging device.
6. Close the discharging device lid.
7. Pull back the contact rocker of the discharge de-
vice and swing it upwards.
Discharging process
1. Slide the locking element back to release the
lifting aid.
2. Press the bellows together and place the suction
pad far left of the middle on the red lid of the
MEDUCORE Easy.
3. Release the bellows. The suction pad will fasten
itself.
4. Pull the lifting aid upwards. This serves to lift the
upper part of the encasing.
5. Slide the locking element forwards to hold the
lifting aid up and in so doing also the upper part
of the encasing.