14
Checking the device
4.10 Checking the condition of the exterior, the equipment and the
accessories
1. Firstly, carry out a visual test of everything.
Requests:
•
Encasing
– Encasing not scratched and without blemishes
– Encasing completely screwed on
– Front film and sticker stuck down firmly and
correctly on the lid
– Labelling OK?
– The loudspeaker opening is free from dirt.
•
Electrode connection
– Undamaged
– In working order
•
Push buttons
– Undamaged
– Working correctly
•
Lid
– Lid not scratched and without blemishes
– Lid is correctly attached, easy to open and
closes firmly
– Functions open, close, switch on are ok.
•
Inner part
– The socket for the electrode plug is free from
dirt.
•
Rechargeable Battery-Pack
– Completely charged
•
Battery-Pack
– Check expiry date (see device sign Battery-
Pack). Insert new Battery-Pack if necessary.
– If the expiry date is not yet reached, then check
the level of charging. Insert battery, open and
close device lid. The green LED must flash; the
yellow LED should not flash.
If these parts have been delivered, check:
• Defibrillation electrodes WM 40116 available.
Check expiry date, if necessary replace.
• Rechargeable Battery-Pack WM 40150
available.
• Battery-Pack WM 40155 available.
• Emergency set WM 15460 available.