Technical data
39
11. Technical data
Device
Dimensions/Environment/Norms
Dimensions L x B x H (in mm incl. handle)
240 x 240 x 93
Weight, empty:
With Rechargeable Battery/Battery-Pack and electrodes:
2.1 kg
2.6 kg
Device class according to MPG and Guidelines 93/42/EEC:
IIb
Operation:
Temperature range:
Without Rechargeable Battery/Battery-Pack and electrodes:
Air humidity:
Air pressure:
0 °C to +50 °C
0 °C to +50 °C
0 % to 95 %
700 to 1060 hPa
Transport/Storage:
0 °C to +50 °C
Temperature range:
max. 2 weeks
-20 °C to +60 °C
Without Rechargeable Battery and electrodes:
-30 °C to +70 °C
Air humidity:
Air pressure:
0 % to 95 %
500 to 1060 hPa
Protective class
IEC 529: IPX4 (protected against sprayed water)
Vibration and knock
DIN-EN 1789:1999
Free fall
EN 60601-1: 1996
Electromagnetic compatibility:
EN 60601-1-2:2001
EN 55011:1998/A1
EN 55014 -1: 2000/A1
EN 61000-4-2:1995/A1/A2
EN 61000--4-3: 1996/A1
EN 61000--4-4.5: 1995
EN 61000--4-6: 1996/A1
EN 61000--4-8: 1993
Norms:
EN 1789, AAMI ANSI DF 39, EN 60601-2-4: 2003, rarely
used
Resuscitation protocol
ERC, AHA; 2005
Self testing
Interval
daily, monthly, when switched on
Time
programmable
Range
battery, electronic, software, charge, shock button, environmental
temperature
Defibrillation electrodes
Condition upon delivery
self sticking once-only electrodes, packed with connecting plug
extruding
Polarisation
not polarized (exchange allowed)
Cable length
125 cm
Electrodes upper surface
every 125 cm
2
service life
30 Months from date of manufacture
Summary of Contents for Homecare MEDUCORE Easy