Repair information and repair instructions
27
8.4 Opening the device
Danger of death! Danger of electric shock! If no discharg-
ing unit is available, work on the MEDUCORE Easy must
be carried out at a protected high-voltage workstation.
1. Take the red device lid
in a safe place.
3. Lie the MEDUCORE Easy with the upper
encasing
facing down
on a non-slip surface.
4. Carefully open the lower encasing device
5. Pull the loudspeaker cable from the printed circuit
board.
6. Pull the red cable plug out of the printed circuit
board.
7. Pull the capacitor out of the lower encasing part
and place it carefully on the main circuit board.
8. Pull the plug belonging to the pad cable
out of the printed circuit board.
9. Pull the blue and black cable plugs out of printed
circuit board.
10. Now you can set aside the lower encasing part
8.5 Exchange the device lids
Remove lid
2. Slide the pins
them.
3. Pull the springs
place.
off.