28
Using the ventilator
EN
position marked by the white triangle, or by switch-
ing off the ventilator.
If audio response is activated, the ventilator confirms
the return to ventilation mode by announcing:
“Mask ventilation mode. Tilt back head, secure mask
tightly” (from serial number 8000 and following re-
pair/servicing/firmware update: “Ventilation pres-
sure limit 20 mbar”).
6.8 Terminating ventilation or Demandflow
Important
Never empty the oxygen
cylinder completely. Re-
turn the cylinder for filling
while it still contains resid-
ual pressure. This prevents
entry of moist atmospheric
air that can cause
corrosion.
1. Check the oxygen supply on the pressure reduc-
er gauge. Always change the cylinder in good
time, e.g. when the pressure is lower than
50 bar, to ensure that oxygen is available for an
adequate period.
2. Close the valve of the oxygen cylinder.
3. Switch off the MEDUMAT Easy. To prevent the
ventilator being switched off unintentionally, ON/
OFF switch must be kept pressed down for at least
2 seconds until the LEDs in the alarm panel
light
up. If audio response is enabled, the ventilator an-
nounces: “Close oxygen cylinder”.